31006 : LBL Labels And Containers Delivered (for Label Broker)

This test concerns the LB (Label Broker) actor. You will need to communicate with the LBL Simulator, in order to simulate the LAB-63 transaction of the LBL Profile.

In this test, the Label Broker notify the effective labeled containers production to the Label Information Provider.
So, the SUT (System Under Test), needs to know the labeling instruction, before to label the tubes and to send a notification message to the LIP for the effective labels printing and labeled containers production.
Two steps are necessary in this test :

  • First, the LBL Simulator will send the labeling instruction to the SUT.
  • Second, the SUT will notify the LBL Simulator for the effective labels printing and labeled containers production.

You can't run this test if you don't passed the "LBL Request Mode For Labeling Instruction" pre-connectathon test.

Instructions


Access the LBL Simulator tool at this location : LBL Simulator
If it is your first time with this tool, please read the user manual : LBL Simulator User Manual
Please be remind that if you are logged in your configurations will be private.

First step :
As your system implements the LB (LB acts as a responder in this part of the test) :

  1. Create or update the configuration corresponding to the SUT.
  2. Go to the "Simulators" menu entry and choose "LIP" page to begin your test.
  3. Select the "LAB-61 Transaction" tab.
  4. In the SUT Configurations drop-down list, select your LB system.
  5. Once you have selected the patient, hit the "Send" button. The LBL Simulator will send the message to your system with the labeling instructions.

Second step :

  1. Go to the "Simulators" menu entry and choose "LIP" page to begin your test.
  2. Select the "LAB-63 Transaction" tab.
  3. In the charset drop-down list, select the desired charset.
  4. Use the Ip Address and the Port linked to this charset to send your notification message to the LBL Simulator for the effective labels printing and labeled containers production.
  5. Don't forget to hit the "Refresh List" button after to have send your message.
  6. Check the simulator has properly received your acknowledgement message.
  7. The messages exchanged between the simulator and the SUT can be found in the message table on the very same page or from the HL7 Message menu entry. Hit the link on the left side  of the raw (first column of the table), to display the Permanent Link to test report for the corresponding HL7 message. (You can access to the hl7v2 report tutorial for more details.) If the validation report status is passed for the message and the response, copy the "test result link" and paste it in Gazelle as the result of this test. For further details, see this tutorial : How enter your pre-connectathon test result in Gazelle (only for the LAW, LCSD and LBL Profiles of the Laboratory Domain which send HL7v2 messages) ?

Evaluation


The validation status must be passed for the two messages of the transaction. The message type must be the right message type, according to IHE.
The Acknowledgment code must be "AA" (MSA-1) in the acknowledgment message.