Details
-
Type: Technical task
-
Status: Closed
-
Priority: High
-
Resolution: Fixed
-
Affects Version/s: 3.3-GA
-
Fix Version/s: 3.3-GA
-
Component/s: Analyzer Manager
-
Labels:None
-
Sprint:Sprint 16, Sprint 17
-
Account:IICC-01 (IICC01)
Description
Action points:
- LAB-27 query SHALL trigger a AWOS broacast event, think about implementing such a mechanism in the simulator
- PID-6/XPN-7 is set to M, SHALL be L
- Usage of SPM-17 depends on value in SPM-11, SHALL take this constraint into account
- Usage of SPM-18 depends on value in SPM-11, SHALL take this constraint into account
- check usage of SAC-13
- ORC-2 SHALL not be used in OML_O33
- Allow user to send a Negative Query response
- Check the content of the SAC segment before sending message to the Analyzer
- Rename ORC-2 and ORC-3 to use AWOS ID and other wording used in TF
- LAB-27 query SHALL trigger a AWOS broacast event, think about implementing such a mechanism in the simulator
- PID-6/XPN-7 is set to M, SHALL be L
- Usage of SPM-17 depends on value in SPM-11, SHALL take this constraint into account
- Usage of SPM-18 depends on value in SPM-11, SHALL take this constraint into account
- check usage of SAC-13
- ORC-2 SHALL not be used in OML_O33
- Allow user to send a Negative Query response
- Check the content of the SAC segment before sending message to the Analyzer
- Rename ORC-2 and ORC-3 to use AWOS ID and other wording used in TF